Dip Switch on the motherboard for a 3000 or 7300 series analyzer
Position Function Normally
1 RS-232 baud rate Off (for 2400 baud rate)
2 RS-232 baud rate Off (for 2400 baud rate)
3 None Off
4 None Off
5 None Off
6 None Off
7 None Off
8 None Off
Dip Switch on the motherboard for a 2000 series analyzer
Position Function Normally
1 RS-232 baud rate Off (for 2400 baud rate)
2 RS-232 baud rate Off (for 2400 baud rate)
3 None Off
4 None Off
5 Reverse polarity of Range 1 Application dependent
6 Reverse polarity of Range 2 Application dependent
7 Reverse polarity of Range 3 Application dependent
8 Reverse polarity of Cal Range Application dependent
Jumpers for the motherboard for a 2000, 3000 or 7300 series analyzer
Position Function
JP1 Selects output voltage of percent of range, should be in position 0
JP2 Selects output voltage of range ID, should be in position 0
JP3 Connects free ADC channel to ground
JP4 Connects free ADC channel to ground
JP5 Connects free ADC channel to ground
JP6 Connects analog and digital grounds together
JP7 Connects analog and digital grounds together


   