A: When a new unit is put in service, it will typically drift fairly steadily for about 7 to 10 days and then finally stabilize out. During this initial period frequently calibrations are necessary, daily at the minimum.
We would recommend that you readjust the pots after several days of purging with O2.
However, once the unit fully stabilizes, as a steady state temperature and gas condition is reached inside the sensor, the drift is typically about 0.025 percent per day (see chart).
So if your required purity is 99.95 per cent, I would recommend a daily calibration at 100%, and a calibration of 95% less frequently, say once per week, once the unit has fully stabilized out.