A: The operator must manually change the pressure setting when backgrounds are changed.
1. Introduce the span gas with an Air background,
2. Increase the fuel 2 psig over the setting specified in the manual.
3. Increase the fuel 3 psig over the setting specified in the manual.
4. Let it run for 20 minutes, then span the analyzer.
5. After the span gas, introduce zero gas (Air) and let it run for 15 minutes to zero the analyzer.
6. Record the air, fuel, and sample pressure for future use.
7. Introduce zero gas (N2) and let it run for 15 to 20 minutes. When the reading stabilizes, select the zero menu and enter the zero value's opposite with the value display.
example: if the reading is -1ppm then enter the zero value is +1ppm
8. Introduce span gas with N2 background and let it run for 15 to 20 minutes. Then increase the sample pressure until the reading displays span gas value. Record the fuel and pressure value for future use. With this method, the operator will have to calibrate zero and span on one background Air and need to change the pressure when backgrounds are changed.